60601-1 Medical DOE EPS 2.0 Level VI Power Supply / Charger, 5v, 2.0A, with integral USB Jack, compliant with 2MOPP, 5000 Meters Elevation and 8 kV Contact/20kV Air Discharge ESD Protection
Information
Model Number
GTM46101-WWVV-X.X-USB
Description
GTM46101-WWVV-X.X-USB, ITE / Medical Power Supply, 60601-1-4th Ed. , Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Blade Options for Q Series Wall Plug-in Power Supplies, Output Rating: 10 Watts, Power rating with convection cooling (W) , 5-5.2V in 0.1V increments, Approvals: South African EAC CAN ICES-3 FCC CE CB 60601-1 2MOPP CB CE cETLus ES60601 cETLus 60601-1 3rd China RoHS Double Insulation IP20 Level VI LPS CB RoHS S-Mark IEC/EN 60601-1 Ukraine WEEE cETLus 60950 RCM CCC India CB EN/IEC 60335-1 PSE
Regulated Switchmode AC-DC Power Supply AC Adaptor
Category
ITE / Medical Power Supply
Input Voltage Range (V)
100-240V~, 50-60 Hz
Max Input Current (A)
0.3A
Max Output Power Rating (W)
10
Output Voltage Range (V)
5-5.2
Efficiency
VI
Ingress Protection
Indoor Use
Size (mm)
41.0 x 71.0 x 31.5
Enclosure
Rating Table
Model Number (Final ordering #/Part No to be defined)
Voltage
Amps (A)
Watts (W)
RFQ
GTM46101-1005-USB*
5
2
10
GTM46101-1306-0.9-USB*
5.1
2.54
12.95
GTM46101-1306-0.8-USB*
5.2
2.5
13
GTM46101-1006-0.75-USB*
5.25
2
10.5
GTM46101-1306-0.7-USB*
5.3
2.45
12.99
GTM46101-1306-0.6-USB*
5.4
2.4
12.96
GTM46101-1006-0.5-USB*
5.5
2
11
GTM46101-0506-0.5-USB*
5.5
0.9
4.95
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS: 1. Input Voltage: Specified 90-264 Vac, Nameplate rated 100-240 Vac 2. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60 Hz 3. Output Regulation: +/- 5% measured at the end of a 1.2 meter USB cordset employing #20 Gauge conductors (RFP RFI-50856) The longest cable GlobTek recommends to be used with this product is USB cord GlobTek PN USBA1M2MICROB20(R), 1.2 Meters for best voltage regulation performance. (RFP RFI-20435-3) 4. Line Voltage Regulation: +/- 3% typical measured at the end of a 1.2 meter USB cordset employing #20 Gauge conductors 5. Output Ripple (Vp-p): 150mV Ripples shall be measured with a 47uF electrolytic capacitor paralleled with a high-frequency 0.1uF ceramic capacitor across the output. The oscilloscope's bandwidth shall be set to 20MHz. 6. Turn-On/Turn-Off Overshoot: 5% maximum, 500 us maximum recovery time for 25% step load. 7. Turn-On Delay: 3000mS MAX at 115VAC Full Load 8. Hold-Up Time: 8mS Min at 100VAC Full Load 9. Inrush Current: No damage. On power when switching on/off repeatedly under any condition. 10. Switching Frequency: 66.5 KHz typical 11. Average Efficiency: 78.7% Min.(CEC&DOE Level VI Compliant), 79.0% Min. (ErP European Commission's Ecodesign Directive (2005/32/EC) / CoC Tier 2 Compliant) 12. Input Power(no load): < 0.1W < 0.1W, < 0.075W (ErP European Commission's Ecodesign Directive (2005/32/EC) / CoC Tier 2 Compliant)
B) PROTECTION 1. Over-Voltage: Output shut down and auto restart 110% to 140% of Vnom 2. Over Temperature Protection: Output shut down and auto restart 3. Short Circuit Protection Output shut down and auto restart 4 Over Load Protection Output shut down and auto restart 110% to 150% of Inom 5. Input Protection: Input line fusing 6. ESD: 8 kV Contact/20kV Air Discharge
C) SAFETY 1. Dielectric Withstand Voltage: 4000Vac or 5656Vdc 10mA 1 minute 2. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so Earth Leakage current is not measured 3. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition) 4. Means of Protection: 2 x MOPP 5. ROHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006
D) EMC 1. Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4 Conducted Emissions: Class B, FCC Part 15, Class B Radiated Emissions: Class B, FCC Part 15, Class B 2. Line Frequency Harmonics EN61000-3-2, Class A 3. Voltage Fluctuations/Flicker EN61000-3-3 4. Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2 Static Discharge Immunity EN61000-4-2, 8kV Contact Discharge, 15kV air discharge Radiated RF Immunity EN61000-4-3, 3V/m 80-2700MHz, 80% 1KHz AM. EFT/Burst Immunity EN61000-4-4, 2kV/100kHz. Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM Power Frequency Magnetic Field Immunity EN61000-4-8, NA Voltage Dip Immunity EN61000-4-11, Criteria
E) OTHER: 1. MTBF: 200,000 hours @ 25°C ambient temperature 2. Operating Temperature: 0°C to 40°C ambient temperature 3. Humidity: 0% to 95% relative humidity 4. Storage Temperature: -40°C to 80°C
F) ENCLOSURE 1. Housing: High impact plastic, 94V0 polycarbonate, non-vented 2. Markings: Label and/or Pad Printed and/or Molded in the case
INPUT CONNECTOR: Q-Socket (below are available blades configurations which are "not included" (unless stated above); can be purchased separately, package with power supply or as a separate "Q-KIT" if specified
01. United States / Canada / Japan NEMA 1-15P/IEC PLUG A[WORKS IN PLUG B] configuration: NA 2 blades, Class II; US/CA/JP P/N: Q-NA(R) 02. Australian AS 3112 configuration: SAA 2 blade/IEC TYPE I, Class II; AU P/N: Q-SAA(R) 03. UK BS 1363 configuration: UK 3 blade with dummy Ground/IEC TYPE G, Class II; GB P/N: Q-UK(R) 04. European CEE 7/16 configuration: Europlug 2 pins/IEC TYPE C [WORKS IN TYPE E&F], Class II; EU P/N: Q-EU(R) 05. Korean KS C8305 configuration: 2 pins/SIMILAR TO IEC TYPE C, Class II; KR P/N: Q-KR(R) 06. Argentina IRAM 2073 configuration: 2 blades/SIMILAR TO IEC TYPE I; AR P/N: Class II Q-AR(R) 07. China GB 2099 configuration: 2 blades/SIMILAR TO TYPE A, Class II; CN P/N:Q-CN(R) 08. India IS 1293 6A/BS546 configuration: 5A, 3 pins with Dummy Ground, Class II/IEC TYPE D; IN P/N: Q-IN(R) 09. IEC320/C18 Inlet, Class II; P/N:Q-C18(R) 10. Brazilian NBR6147 configuration: 2 pins, Class II;SIMILAR TO IEC TYPE C BR P/N: Q-BR(R) 11. South Africa SABS164-1, 3 round prongs, Class II + dummy ground, IEC TYPE M P/N: Q-SANS164-1-16A(R) 12. South Africa SABS164-4, 3 round prongs with a notched prong @ 0º, Class II + dummy ground, IEC TYPE M Red, P/N: Q-SANS164-4L-16A(R) 13. South Africa SABS164-4, 3 round prongs with a notched prong @ -53º, Class II + dummy ground, Blue, IEC TYPE M, P/N: Q-SANS164-4C-16A(R) 14. South Africa SABS164-4, 3 round prongs with a notched prong @ +53º, Class II + dummy ground, Black,IEC TYPE M P/N: Q-SANS164-4R-16A(R)
Contact GlobTek for your specific requirements or custom solutions.
Approvals
Logo
Description
Bureau Of Indian Standards for GTM46101-1005-USB (5V only)
Book 60335 up to 36 Volts
Book 60601-3
No Logo Applicable
CB report IEC60601-1 2005 A1+C1+C2 2016-2-4 and or EN 60601-1:2006 3.1rd Edition 2xMOPP
No Logo Applicable
CB for IEC 62368-1:2014 (Second Edition)
No Logo
CB Report IEC60950-1:2005 (2nd Edition), Am 1: 2009 + Am 2:2013
No Logo Applicable
CB to 60335-1 5th Edition (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) CB Report/ Certificate IEC 60335-1:2010 Up to 36 volts only
CCC to GB4943.1-2011 GB9254-2008 GB17625.1-2012 with Non-Tropical and Altitude above 2000m approval
CE Certification
CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2013(Ed4) EN 55024.2010
CE Mark: tested to comply with EN60601-1-2:2013 (4th Ed), EN61000-3-2, EN61000-3-3, EN50082-1, including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11, EN61000-1-2-2001
EMI: EN55011 CLASS B and FCC Part 15 B
Conforms to ANSI/AAMI ES 60601-1:2005 Cert.to CAN/CSA Std. C22.2
cETLus ANSI_AAMI ES60601-1_2005
Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
[AAMI ES60601-1:2005 +A1]Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance [CSA C22.2#60601-1:2014
Information Technology Equipment Safety Part 1: General Requirements (UL 60950-1 Issued: 2007/03/27
Ed: 2 Rev: 2014/10/14)
Information Technology Equipment Safety Part 1: General Requirements (CSA C22.2 No. 60950-1 Issued:2007/03/27 Ed: 2 (R2012) Amd. 1
CHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
Declaration # ???? N RU ?-US.??75.?.01052 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
Compliance of this power supply with FCC Part 15, Class B has been demonstrated with a standard output load. The FCC law stipulates that system-level testing is required to demonstrate compliance with the FCC emission limits with the actual system load.
Fuse 60335 T1.0A 250VAC Up to 36 volts
CAN ICES-3(B)/NMB-3(B)
Compliance of this PSU with Industry Canada, Class B demonstrated with a standard output load. The ICES law stipulates that system-level testing is required to demonstrate compliance with the ICES-3 emission limits with the actual system load.
Indoor Use Only - Mark is on the label or Molded in the case
IP20
Ingress Protection: IP20 to IEC60529:2001 Protected against solids objects over 12mm and No Protection
JAPAN TUV R-PSE, Cert. No. JD 50315169, to J60950-1(H26) , J55022(H22),[15V or less]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
LPS
Limited Power Source
No Logo
SANS IEC 60950-1
RCM certificate SAA-152394-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm)
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
IEC62368 Power Supply Adapter meets DOE Level VI/EPS 2.0/ErP Tier2 requirements and is certified to Medical and ITE standards. This Power Supply / Charger offers Integral USB Jack and Interchangeable International Blades, GTM46101-1005-USB